A bad workman blames his tools?
Related Experiment Video
Updated: Apr 22, 2026

In Situ Soil Moisture Sensors in Undisturbed Soils
Published on: November 18, 2022
Heather Jean Edwards1, Huda Al-Kutubi2, Nia Rathbone2
1Department of Medicine, Cardiff University, Cardiff, UK.
Abstract:
The ECG is a test that is used frequently in the acute setting. It has a significant impact on decisions regarding patient discharge and further investigations. On a single day in the ambulatory emergency care setting two patients presented with chest pain. The ECG findings were abnormal, but also out of context with the clinical findings. On close inspection of the ECG machine it was identified that although all leads attached to the patient were in the correct position, the two cables connecting the leads to the machine had been reversed. Had the error not been discovered promptly there was the potential that further, more harmful investigations would have been performed. These cases highlight that although the ECG is a simple and non-invasive investigation it should be an adjunct to clinical working diagnosis.