Combined Raman microspectrometer and shearforce regulated SECM for corrosion and self-healing analysis
Mathieu Etienne1, Manuel Dossot, Jérôme Grausem
1CNRS and Université de Lorraine, Laboratoire de Chimie Physique et Microbiologie pour l'Environnement, UMR 7564 , 405 rue de Vandoeuvre, F-54600 Villers-lès-Nancy, France.
Abstract:
Shearforce regulated scanning electrochemical microscopy (SECM) has been associated with Raman microspectrometry in order to perform combined electrochemical and spectrochemical analysis on reactive interfaces. The interest of the method was evaluated by analyzing local corrosion phenomena in damaged Zn(Mg, Al) self-healing coatings deposited on steel. Despite the high aspect ratio of the analyzed sample displaying here more than a 50 μm depth profile, the optimized setup allowed (1) precise electrode positioning with the help of shearforce detection, (2) electrochemical measurement at a constant distance from the sample surface, and (3) local chemical analysis of the solid surface by confocal Raman microspectroscopy performed at a constant focal distance from the sample. All in all, this new setup allows one to approach the detailed reactivity involved in defective metal samples.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...