Simulated flying altitude and performance of continuous positive airway pressure devices
Maria Sehlin1, Helge Brändström, Ola Winsö
1Department of Surgical and Perioperative Sciences, Anesthesiology and Intensive Care Medicine, Umeå University, Umeå, Sweden.
Introduction:
Continuous positive airway pressure (CPAP) is used in air ambulances to treat patients with impaired oxygenation. Differences in mechanical principles between CPAP devices may affect their performance at different ambient air pressures, as will occur in an air ambulance during flight.
Methods:
Two different CPAP systems, a threshold resistor device and a flow resistor device, at settings of 5 and 10 cm H₂O were examined. Static pressure, static airflow, and pressure during simulated breathing were measured at ground level and at three different altitudes [2400 m (7874 ft), 3000 m (9843 ft), and 10,700 m (35,105 ft)].
Results:
When altitude increased, the performance of the two CPAP systems differed during both static and simulated breathing pressure measurements. With the threshold resistor CPAP, measured pressure levels were close to the preset CPAP level. Static pressure decreased 0.71 ± 0.35 cm H₂O at CPAP 10 cm H₂O compared to ground level and 35,105 ft (10,700 m). With the flow resistor CPAP, as the altitude increased, CPAP produced pressure levels increased. At 35,105 ft (10,700 m), the increase was 5.13 ± 0.33 cm H₂O at CPAP 10 cm H₂O.
Discussion:
The velocity of airflow through the flow resistor CPAP device is strongly influenced by reduced ambient air pressure, leading to a higher delivered CPAP effect than the preset CPAP level. Threshold resistor CPAP devices seem to have robust performance regardless of altitude. Thus, the threshold resistor CPAP device is probably more appropriate for CPAP treatment in an air ambulance cabin, where ambient pressure will vary during patient transport.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...