Within-clinic reagent strip lipid measurement
H A Neil1, D M Cassidy, M F Laker
1Department of Medicine, University of Newcastle upon Tyne, UK.
Diabetic Medicine : a Journal of the British Diabetic Association
|December 1, 1989
Summary
The Reflotron dry chemistry analyzer shows wide agreement limits for cholesterol and triglycerides compared to lab methods. Clinicians must consider these differences for accurate hyperlipidemia screening and patient monitoring.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Biomedical Instrumentation
Background:
- Dry chemistry analyzers offer point-of-care testing advantages.
- Accurate lipid measurements are crucial for diagnosing and managing hyperlipidemia.
Purpose of the Study:
- To assess the agreement between the Reflotron dry chemistry analyzer and standard laboratory methods for total cholesterol and triglycerides.
- To evaluate between-method variability and its clinical implications.
Main Methods:
- Paired comparison of replicate measurements from venous blood samples.
- Analysis of 105 cholesterol and 90 triglyceride specimens.
- Assessment of mean differences, limits of agreement, and repeatability coefficients.
Main Results:
- No significant mean difference for cholesterol, but wide limits of agreement (-0.87 to 0.90 mmol/L) indicate potential clinical discrepancies.
- Poor repeatability for both methods limited agreement.
- Significant systematic difference for triglycerides, with Reflotron reading lower (0.30 mmol/L less) than laboratory methods, and limits of agreement of -0.09 to 0.70 mmol/L.
Conclusions:
- The Reflotron analyzer exhibits clinically significant variability for cholesterol and triglyceride measurements compared to laboratory methods.
- Clinicians should be aware of these agreement limits when using the Reflotron for hyperlipidemia screening or monitoring diabetic patients.
- Poor repeatability in both analytical systems contributes to the observed discrepancies.