Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state camera
Amir Pourmoghaddas1, Karen Vanderwerf, Terrence D Ruddy
1Department of Physics, Carleton University, Ottawa, ON, Canada, apour@ottawaheart.ca.
Summary
Dual-energy-window scatter correction (DEW-SC) is feasible for cadmium-zinc-telluride (CZT) cardiac cameras, producing results comparable to conventional SPECT cameras. This technique accurately corrects myocardial perfusion images without introducing significant bias.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Cardiology
Background:
- Photon attenuation and scatter degrade image quality in conventional gamma cameras.
- Evaluation of scatter correction (SC) for novel solid-state cardiac cameras is limited.
- Cadmium-zinc-telluride (CZT) detectors offer potential advancements in cardiac imaging.
Purpose of the Study:
- To assess the accuracy of dual-energy-window (DEW) scatter correction (SC) on CZT cardiac cameras.
- To compare DEW-SC images from a CZT camera with those from a conventional SPECT camera.
- To evaluate the impact of SC on myocardial perfusion imaging using Tc-99m-tetrofosmin.
Main Methods:
- A modified DEW SC method was developed for CZT detectors, accounting for primary photons in the lower energy window.
- Phantom experiments were conducted to estimate DEW correction parameters.
- Clinical data from 108 patients were analyzed, comparing images reconstructed with no correction (NC), attenuation correction (AC), and AC with DEW-SC from both CZT and conventional SPECT cameras.
Main Results:
- Excellent correlation (r ≥ 0.94) was observed between summed stress/rest scores (SSS/SRS) from both cameras.
- Attenuation correction (AC) reduced the mean difference between cameras for SSS/SRS compared to NC.
- DEW-SC on the CZT camera reduced bias observed without SC, and AC/SC images showed no significant difference between cameras, despite increased noise.
Conclusions:
- Dual-energy-window scatter correction (DEW-SC) is feasible for CZT cardiac cameras.
- DEW-SC produces myocardial perfusion images comparable to conventional SPECT cameras.
- While SC increases noise on CZT images, it does not introduce bias relative to conventional systems.