Related Experiment Video
Updated: Jan 30, 2026

Measuring the Behavioral Effects of Intraocular Scatter
Published on: February 18, 2021
Optical bench performance of AcrySof(®) IQ ReSTOR(®), AT LISA(®) tri, and FineVision(®) intraocular lenses
Daniel Carson1, Warren E Hill2, Xin Hong1
1Alcon Research, Ltd., Fort Worth, TX, USA.
Purpose:
To compare the resolution and optical quality of the ReSTOR(®) +3.0 D and ReSTOR +2.5 D multifocal intraocular lenses (IOLs) with the AT LISA(®) tri and FineVision(®) trifocal IOLs.
Methods:
Resolution, image quality, and photic phenomena were evaluated in the AcrySof(®) IQ ReSTOR +3.0 D and +2.5 D multifocal IOLs and compared with the AT LISA tri 839MP and FineVision Micro F12 trifocal IOLs, using a Badal optometer and a Snellen visual acuity chart. Simulated headlight images were obtained using a modulation transfer function (MTF) bench and a 50 μm pinhole target. MTF values, using vertical and horizontal slits, were determined at far, intermediate, and near distances.
Results:
Resolution at 20/40 Snellen visual acuity equivalence was attainable over nearly the entire viewing distance range with the AT LISA tri and FineVision IOLs, but background shadows were more prominent with the AT LISA tri and FineVision IOLs than with the ReSTOR IOLs. Distance MTF peaks at 20/20 Snellen-equivalent spatial frequency were greatest for ReSTOR +2.5 D and ReSTOR +3.0 D IOLs. The near MTF peak occurred at 53 cm with ReSTOR +2.5 D and had a 20/20 Snellen-equivalent value that was lower than the near peaks of the other models but higher than the intermediate foci of the trifocal IOLs.
Conclusion:
AT LISA tri and FineVision trifocal IOLs achieved a useful third focus for intermediate vision but were associated with increased background halos and reduced distance visual quality compared with ReSTOR +2.5 D and +3.0 D multifocal IOLs.
More Related Videos
10:10Ultrahigh Resolution Mouse Optical Coherence Tomography to Aid Intraocular Injection in Retinal Gene Therapy Research
Published on: November 2, 2018
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Sources of Self-Esteem II: Performance Feedback
Restorative Care
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Pathophysiology of Cardiac Performance
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation