Fluid dynamics and intraocular pressure using venturi phacoemulsification machine in dogs ex vivo
Related Experiment Video
Updated: Apr 21, 2026

Full-Circle Cauterization of Limbal Vascular Plexus for Surgically Induced Glaucoma in Rodents
Published on: February 15, 2022
Seonmi Kang1, Sangwan Park1, Hyunwoo Noh1
1Department of Veterinary Clinical Sciences, College of Veterinary Medicine and Research Institute for Veterinary Science, Seoul National University, 1 Gwanak-ro, Gwanak-gu, Seoul, 151-742, Korea.
Objective:
To optimize fluid dynamics through measuring intraocular pressure (IOP) in a venturi phacoemulsification machine in dogs.
Procedures:
In step I, flow and IOP of the test chamber were measured using a pressure transducer with the bottle height (BH) set at 50, 70, 100, and 120 cm and the vacuum from 30 to 450 mmHg. A 19-gauge phaco and a 0.3-mm irrigation/aspiration (I/A) handpiece were used. In step II, flow and IOP were measured in an enucleated canine eye with a 3.0- and a 3.2-mm clear corneal incision (CCI), respectively. IOP was measured using the pressure transducer at a 30-mmHg vacuum to allow corneal deformation, to define the fluidic parameters for preventing surge.
Results:
Flow was directly proportional to the BH and vacuum, whereas IOP was directly proportional to the BH and inversely to the vacuum. Flow with an I/A handpiece was significantly less than with a phaco handpiece, explaining why IOP with an I/A handpiece was significantly higher than with a phaco handpiece. With the I/A handpiece, vacuum parameters less than 450 mmHg did not result in corneal deformation. IOP with a 3.2-mm CCI was significantly lower than with a 3.0-mm CCI, with the 3.2-mm flow being greater than the 3.0-mm flow.
Conclusions:
BH during the I/A stage could be reduced to avoid unnecessary stress on the canine eye when using a venturi system. Although phacoemulsification with a 3.2-mm CCI could induce lower IOP, a 3.0-mm CCI might lessen the irrigation flow stress on the eye.