Analysis of imprecision in incurred sample reanalysis for small molecules
Related Experiment Video
Updated: Apr 21, 2026

Selected Reaction Monitoring Mass Spectrometry for Absolute Protein Quantification
Published on: August 17, 2015
Sriram Subramaniam1, Devvrat Patel, Barbara M Davit
1Division of Bioequivalence III, Office of Bioequivalence, Office of Generic Drugs, Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland, 20993, USA, sriram.subramaniam@fda.hhs.gov.
Abstract:
Over the years, incurred sample (IS) reanalysis (ISR) has become a tool to confirm the reliability of bioanalytical measurements. The recommendation for ISR acceptance criterion for small molecules is at least 67% of ISR samples that have reanalyzed concentrations within 20% of their original concentrations when normalized to their means. To understand the relevance of the ISR acceptance criterion and sample size requirements, simulated ISR studies evaluated the probability of ISR studies passing the acceptance criterion (ISR pass rate) as a function of IS imprecision and sample size. When IS imprecision (percent coefficient of variation: %CV) is low (≤ 10 or 1-10% CV), high ISR pass rate (≥ 99%) is attained with <50 samples. At intermediate IS imprecision (e.g., 12% CV or 7-12% CV range), 80-160 samples are required for a high ISR pass rate. When IS imprecision is at the higher end of the acceptance limit, ISR pass rate decreases significantly, and increasing sample size fails to achieve high ISR pass rate. The effect of systematic bias (e.g., instability, interconversion) on ISR pass rate is strongly dependent on sample size at intermediate IS imprecision. The results provide an understanding of the effect of IS imprecision on ISR pass rates and a framework for selection of ISR sample sizes.
More Related Videos
08:00Sample Preparation and Relative Quantitation using Reductive Methylation of Amines for Peptidomics Studies
Published on: November 4, 2021
08:43PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis
Published on: May 11, 2017
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Analyzers: Overview
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...