Related Experiment Video
Updated: Apr 21, 2026

Macro-Rheology Characterization of Gill Raker Mucus in the Silver Carp, Hypophthalmichthys molitrix
Published on: July 10, 2020
Design and development of a rheometer for biological fluids of limited availability
A Scorza1, L Battista2, S Silvestri3
1Department of Engineering, Roma TRE University, via della Vasca Navale 79/81, 00146 Roma, Italy.
Abstract:
From studies on the dynamic characterization of human bones, it is noticed that reference data on the viscous behavior of the bone marrow are quite poor. Dependently from marrow limited availability and its opacity, we have not been able to retrieve a tool of appropriate characteristics able to measure bone marrow viscosity. Therefore, principal techniques for the viscosity measurement have been preliminarily examined, and a device suitable for viscosity measurements of biological fluids has been realized. In particular, a rotational rheometer has been developed: it is a coaxial cylinders system, where the fluid flows dragged by the inner cylinder. The device is an absolute rheometer, that is, particularly useful as nowadays it is not known the classification of the bone as far as it concerns its viscous behavior. In this work a preliminary evaluation of the metrological characteristics of the measurement system has been carried out and its main metrological performances have been evaluated.