Related Experiment Video
Updated: Apr 21, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
The upstream boundary condition influences the leaflet opening dynamics in the numerical FSI simulation of an aortic
Sebastiaan Annerel1, Joris Degroote, Tom Claessens
1Department of Flow, Heat and Combustion Mechanics, Ghent University, Sint-Pietersnieuwstraat 41, B-9000, Ghent, Belgium. Sebastiaan.Annerel@UGent.be.
Abstract:
In this paper, the influence of the upstream boundary condition in the numerical simulation of an aortic bileaflet mechanical heart valve (BMHV) is studied. Three three-dimensional cases with different upstream boundary conditions are compared. The first case consists of a rigid straight tube with a velocity profile at its inlet. In the second case, the upstream geometry is a contracting left ventricle (LV), positioned symmetrically with respect to the valve. In the last case, the LV is positioned asymmetrical with respect to the valve. The cases are used to simulate the same three-dimensional BMHV. The change in time of the LV volume is calculated such that the flow rate through the valve is identical in each case. The opening dynamics of the BMHV are modelled using fluid-structure interaction. The simulations show that differences occur in the leaflet movement of the three cases. In particular, with the asymmetric LV, one of the leaflets impacts the blocking mechanism at its open position with a 34% higher velocity than when using the velocity profile, and with an 88% higher velocity than in the symmetric LV case. Therefore, when simulating such an impact, the upstream boundary condition needs to be chosen carefully.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Fluid Pressure over Curved Plate of Constant Width
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Fluid Pressure over Flat Plate of Constant Width
The resultant force...
Design Example: Forces in Sluice Gate
Key variables in...
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in...