Flat-panel cone-beam CT-guided radiofrequency ablation of very small (≤ 1.5 cm) liver tumors: technical note on a
Roberto Luigi Cazzato1, Xavier Buy, Nicolas Alberti
1Department of Radiology, Institut Bergonié 229 Cours de l'Argonne, 33000, Bordeaux, France, r.cazzato@unicampus.it.
Purpose:
The aim of the present study was to investigate the technical feasibility of flat-panel cone-beam CT (CBCT)-guided radiofrequency ablation (RFA) of very small (<1.5 cm) liver tumors.
Materials And Methods:
Patients included were candidates for hepatic percutaneous RFA as they had single biopsy-proven hepatic tumors sized ≤ 1.5 cm and poorly defined on ultrasonography. Following apnea induction, unenhanced CBCT scans were acquired and used to deploy the RF electrode with the aid of a virtual navigation system. If the tumor was not clearly identified on the unenhanced CBCT scan, a right retrograde arterial femoral access was established to carry out hepatic angiography and localize the tumor. Patients' lesions and procedural variables were recorded and analyzed.
Results:
Three patients (2 male and 1 female), aged 68, 76, and 87 years were included; 3 lesions (2 hepato-cellular carcinoma and 1 metastasis from colorectal cancer) were treated. One patient required hepatic angiography. Cycles of apnea used to acquire CBCT images and to deploy the electrode lasted <120 s. Mean fluoroscopic time needed to deploy the electrode was 36.6 ± 5.7 min. Mean overall procedural time was 66.0 ± 22.9 min. No peri- or post-procedural complications were noted. No cases of incomplete ablation were noted at 1-month follow-up.
Conclusion:
Percutaneous CBCT-guided liver RFA with or without arterial hepatic angiography is technically feasible.