Comparison of two multifocal intraocular lens designs that differ only in near add
Purpose:
To evaluate monocular functional outcomes after the implantation of the AcrySof ReSTOR SN6AD2 intraocular lens (IOL) (+2.5 diopters [D] near add) (Alcon Laboratories, Inc., Fort Worth, TX) and the AcrySof ReSTOR SN6AD1 IOL (+3.0 D near add) (Alcon Laboratories, Inc.).
Methods:
This prospective, comparative, nonrandomized single-blind observational study comprised 62 eyes of 62 patients that underwent phacoemulsification and implantation of a multifocal IOL: SN6AD2 (31 eyes) (+2.5 group) and SN6AD1 (31 eyes) (+3.0 group). Twelve months after surgery, monocular near (30 and 40 cm), intermediate (50, 60, and 70 cm), and distance (4 m) visual acuity were evaluated with the internal root mean square and modulation transfer function. Both parameters were evaluated at the 4- and 6-mm pupil sizes.
Results:
No statistical differences at 4 m were found between the groups. The +2.5 group obtained better performances at all intermediate distances (50 cm, 0.23 ± 0.14 vs 0.32 ± 0.13; 60 and 70 cm, 0.21 ± 0.10 vs 0.41 ± 0.14 and 0.24 ± 0.10 vs 0.53 ± 0.17, respectively), whereas the near visual acuity was better for the +3.0 group (30 and 40 cm, 0.36 ± 0.18 vs 0.14 ± 0.09 and 0.36 ± 0.19 vs 0.17 ± 0.07, respectively). The root mean square was lower for the +2.5 group compared to the +3.0 group, whereas the modulation transfer function showed overlapping results between the two models.
Conclusions:
Both IOL models showed good results in distance vision; the +2.5 D IOL seemed to provide better intermediate vision than the +3.0 D IOL. For near vision, the +3.0 D model performed better than the +2.5 D model.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...