Related Experiment Video
Updated: Apr 21, 2026

Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
Grueneisen relaxation photoacoustic microscopy
Lidai Wang1, Chi Zhang1, Lihong V Wang1
1Optical Imaging Laboratory, Department of Biomedical Engineering, Washington University in St. Louis, Campus Box 1097, One Brookings Drive, St. Louis, Missouri 63130-4899, USA.
Abstract:
The temperature-dependent property of the Grueneisen parameter has been employed in photoacoustic imaging mainly to measure tissue temperature. Here we explore this property using a different approach and develop Grueneisen relaxation photoacoustic microscopy (GR-PAM), a technique that images nonradiative absorption with confocal optical resolution. GR-PAM sequentially delivers two identical laser pulses with a microsecond-scale time delay. The first laser pulse generates a photoacoustic signal and thermally tags the in-focus absorbers. When the second laser pulse excites the tagged absorbers within the thermal relaxation time, a photoacoustic signal stronger than the first one is produced, owing to the temperature dependence of the Grueneisen parameter. GR-PAM detects the amplitude difference between the two colocated photoacoustic signals, confocally imaging the nonradiative absorption. We greatly improved axial resolution from 45 μm to 2.3 μm and, at the same time, slightly improved lateral resolution from 0.63 μm to 0.41 μm. In addition, the optical sectioning capability facilitates the measurement of the absolute absorption coefficient without fluence calibration.