Self-repositioning of an embolized patent ductus arteriosus device--a nightmare turned into sweet dreams
Related Experiment Video
Updated: Apr 21, 2026

Closure of a Patent Foramen Ovale PFO: An Intervention Sequence
Published on: December 23, 2022
Venkateshwaran Subramanian1, Krishnamoorthy Kavassery Mahadevan, Sivasankaran Sivasubramonian
1Department of cardiology, Sreechitra Tirunal institute for medical sciences and technology, Thiruvananthapuram, Kerala, India.
Abstract:
A 7-month-old boy was admitted for the device closure of symptomatic patent ductus arteriosus (PDA) with moderate pulmonary hypertension. The PDA measured 4.2 mm with adequate ampulla. It was closed with an 8-6 mm Heart R device from pulmonary artery (PA) end. Post extubation, the device embolized to proximal descending thoracic aorta just distal to PDA ampulla. While attempting to snare from the venous side, the device self-repositioned to PDA. It was stable thereafter and patient was discharged after 2 days. We report a complication, which got self-corrected.