Technical and radiological image quality comparison of different liquid crystal displays for radiology
Francina Em Dams1, K Y Esther Leung2, Pieter Hm van der Valk1
1Department of Radiology, Albert Schweitzer Hospital, Dordrecht, The Netherlands.
Background:
To inform cost-effective decisions in purchasing new medical liquid crystal displays, we compared the image quality in displays made by three manufacturers.
Methods:
We recruited 19 radiologists and residents to compare the image quality of four liquid crystal displays, including 3-megapixel Barco(®), Eizo(®), and NEC(®) displays and a 6-megapixel Barco display. The evaluators were blinded to the manufacturers' names. Technical assessments were based on acceptance criteria and test patterns proposed by the American Association of Physicists in Medicine. Radiological assessments were performed on images from the American Association of Physicists in Medicine Task Group 18. They included X-ray images of the thorax, knee, and breast, a computed tomographic image of the thorax, and a magnetic resonance image of the brain. Image quality was scored on an analog scale (range 0-10). Statistical analysis was performed with repeated-measures analysis of variance.
Results:
The Barco 3-megapixel display passed all acceptance criteria. The Eizo and NEC displays passed the acceptance criteria, except for the darkest pixel value in the grayscale display function. The Barco 6-megapixel display failed criteria for the maximum luminance response and the veiling glare. Mean radiological assessment scores were 7.8±1.1 (Barco 3-megapixel), 7.8±1.2 (Eizo), 8.1±1.0 (NEC), and 8.1±1.0 (Barco 6-megapixel). No significant differences were found between displays.
Conclusion:
According to the tested criteria, all the displays had comparable image quality; however, there was a three-fold difference in price between the most and least expensive displays.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Radiological Investigation I: X-ray and CT
Three-Dimensional Microscopy in Microbiology
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography