Raman difference spectroscopy: a non-invasive method for identification of oral squamous cell carcinoma
Knipfer Christian1, Motz Johanna2, Adler Werner3
1Department of Oral and Maxillofacial Surgery, Friedrich-Alexander Universität Erlangen-Nürnberg (FAU), Erlangen, Germany ; Erlangen Graduate School in Advanced Optical Technologies (SAOT), Friedrich-Alexander Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Abstract:
The feasibility of shifted-excitation Raman difference spectroscopy (SERDS) as a label-free and non-invasive technique for an objective diagnosis of oral cancer (OSCC) was investigated by analyzing 12 ex vivo OSCC samples. 72 mean SERDS spectra from each three physiological tissue points and pathological lesions were correlated with the histo-pathological diagnosis. Principal component analysis (PCA) and linear discriminant analysis (LDA) showed excellent results with an area under the curve of 94.5% and a classification error of 9.7% (sensitivity: 86.1%; specificity: 94.4%). The SERDS Raman spectra of malignant and benignant tissues were discriminable with respect to the spectral features of proteins, lipids and nucleic acids. The presented method is capable of a highly accurate identification of OSCC. These findings suggest a high validity and reproducibility of SERDS combined with PCA and LDA analysis regarding oral cancer tissue.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...