Background:
Transcutaneous carbon dioxide (PtcCO2 ) monitoring is rarely used in the acute hospital setting, where serial samples of arterial blood are instead taken to measure carbon dioxide tension (PaCO2 ). In this pilot observational study, we assessed the potential of PtcCO2 monitoring to calculate pH and guide management of acute noninvasive ventilation (NIV).
Methods:
Ten subjects with acute hypercapnic respiratory failure were recruited. All had arterial lines placed to guide acute NIV. PtcCO2 was monitored for 12 h (TOSCA TCM4) and compared with PaCO2 . Noninvasive transcutaneous pH was determined from PtcCO2 and calculated bicarbonate and then compared with true arterial pH. Agreements between PCO2 and pH methods were assessed using Bland-Altman analysis of limits of agreement and Pearson correlation coefficients. Hypothetical adjustments to acute NIV settings were based on transcutaneous data alone and evaluated in comparison with true management. Pain scores for each method were compared using the Wilcoxon signed-rank test.
Results:
PCO2 time trends were concordant. Mean PCO2 bias was -2.33 (95% limits of agreement of -9.60 to 5.03) mm Hg, and r = 0.89 (P < .001). Mean pH bias was 0.012 (95% limits of agreement of -0.070 to 0.094), and r = 0.84 (P < .001). Hypothetical clinical decisions based on transcutaneous data alone matched true management on 85% of 34 occasions. Initiation of transcutaneous monitoring was less painful than the arterial equivalent (P = .008).
Conclusions:
This pilot study demonstrates that PtcCO2 monitoring provides a continuous and reliable trend and also allows pH prediction. This patient-friendly approach is a promising alternative to repeated arterial blood gas sampling in patients requiring NIV for acute hypercapnic respiratory failure.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-III
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...