Comparability of hemoglobin A1c level measured in capillary versus venous blood sample applying two point-of-care

Tahereh Keramati1, Farideh Razi2, Ali Tootee2

  • 1Endocrinology and Metabolism Research Center, Endocrinology and Metabolism Clinical Sciences Institute, Tehran University of Medical Sciences, Tehran, Iran.

Insights

Capillary blood hemoglobin A1c measurement is a valid alternative to venous blood testing for monitoring diabetes control. This study confirms its accuracy in differentiating between well-controlled and poorly-controlled blood glucose levels.

Area of Science:

  • Clinical Chemistry
  • Endocrinology
  • Diabetes Management

Background:

  • Hemoglobin A1c (HbA1c) is a key biomarker for long-term glycemic control in diabetes.
  • Venous blood sampling for HbA1c is standard, but capillary blood offers a less invasive alternative.
  • Evaluating the accuracy of capillary HbA1c against venous HbA1c is crucial for clinical practice.

Purpose of the Study:

  • To assess the validity of capillary blood hemoglobin A1c (HbA1c) measurements compared to venous blood HbA1c.
  • To determine the reliability of capillary blood HbA1c for distinguishing between different levels of glycemic control.

Main Methods:

  • A cross-sectional study involving 45 Iranian diabetic patients.
  • Simultaneous venous and capillary blood samples were collected for HbA1c analysis.
  • HbA1c levels were measured using two NGSP-certified systems: CERA STAT 2000 and NycoCard Reader II.

Main Results:

  • Strong correlations were found between venous and capillary HbA1c levels (Pearson's concordance correlation coefficients of 0.96 and 0.94).
  • Both CERA STAT 2000 (AUC=0.991) and NycoCard Reader II (AUC=0.935) assays showed high accuracy in differentiating glycemic control levels using capillary blood.
  • While CERA STAT 2000 showed no significant difference between venous and capillary means (p=0.590), NycoCard Reader II showed a significant difference (p=0.007).

Conclusions:

  • Capillary blood hemoglobin A1c measurement is a highly valuable tool for differentiating between poorly-controlled and well-controlled blood glucose levels.
  • The findings support the use of capillary blood HbA1c as a reliable alternative to venous blood testing in diabetes management.
Abstract

Related Concept Videos

Capillary Electrophoresis: Instrumentation01:20

Capillary Electrophoresis: Instrumentation

Capillary electrophoresis instrumentation typically consists of several key components. A high-voltage power supply generates the electric field necessary for the separation by connecting to an anode (the positively charged electrode) and a cathode (the negatively charged electrode) located in buffer reservoirs at each end of the capillary tube. The system includes a sample vial, a fused silica capillary tube coated with polyimide for mechanical strength through which the sample components...
1.8K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
2.1K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
4.0K