Radiation dose reduction at MDCT with iterative reconstruction for prenatal diagnosis of skeletal dysplasia:
Chihiro Tani1, Yoshinori Funama, Chikako Fujioka
11 Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minamiku, Hiroshima 734-8551, Japan.
Objective:
The purpose of this study was to investigate to what degree the radiation dose can be reduced without affecting the ability to evaluate normal fetal bones at MDCT with iterative reconstruction.
Materials And Methods:
Fifteen normal fetal specimens immersed in containers (30- and 35-cm diameter) were scanned with a 64-MDCT scanner, with tube voltage of 100 kVp and tube current of 600, 300, 150, 100, and 50 mA. Images were subjected to adaptive statistical iterative reconstruction (ASIR). The fetal dose was measured using glass dosimeters. We calculated the relative ratio of the dose at 600 mA. Image quality was evaluated on maximum-intensity-projection and volume-rendering images. Two radiologists recorded the visualization scores of five regions. Images at 600 mA were considered to be standard.
Results:
With the 30-cm-diameter container, the fetal dose was 10.15 mGy (relative ratio, 100%) at a tube current of 600, 51% at 300, 25% at 150, 17% at 100, and 9% at 50 mA. With the 35-cm-diameter container the fetal dose was 10.01 mGy (relative ratio, 100%) at 600, 47% at 300, 24% at 150, 17% at 100, and 8% at 50 mA. Visual evaluation showed that in both containers, with ASIR 90%, there was a statistically significant difference between 50-and 600-mA images (p<0.01) but not between 600-mA images and those acquired at 100, 150, and 300 mA (p=0.08-1.00).
Conclusion:
The fetal radiation dose for the evaluation of normal fetal bones can be reduced by 83% with ASIR 90%.
Insights
Radiation dose for fetal bone evaluation using MDCT can be significantly reduced by 83% with adaptive statistical iterative reconstruction (ASIR) 90%, without compromising image quality for diagnostic purposes.
Area of Science:
- Medical Imaging
- Radiology
- Pediatric Imaging
Background:
- Multidetector computed tomography (MDCT) is crucial for fetal bone assessment.
- Optimizing radiation dose is essential in pediatric and prenatal imaging to minimize risks.
Purpose of the Study:
- To determine the extent of radiation dose reduction possible in fetal MDCT scans.
- To evaluate if reduced doses impact the diagnostic quality of normal fetal bone visualization using iterative reconstruction.
Main Methods:
- Fifteen fetal specimens were scanned using a 64-MDCT scanner at varying tube currents (50-600 mA) and 100 kVp.
- Images were reconstructed using adaptive statistical iterative reconstruction (ASIR) 90%.
- Fetal radiation dose was measured, and image quality was assessed by two radiologists.
Main Results:
- Radiation dose was reduced by up to 91% compared to the 600 mA standard.
- No statistically significant difference in image quality was found between standard dose (600 mA) and doses reduced to 100, 150, and 300 mA when using ASIR 90%.
- A significant difference was observed only between the lowest dose (50 mA) and the standard dose.
Conclusions:
- Fetal radiation dose can be reduced by 83% using ASIR 90% while maintaining diagnostic image quality for fetal bone evaluation.
- Iterative reconstruction techniques like ASIR allow for substantial radiation dose reduction in fetal MDCT.
- This facilitates safer and more efficient prenatal imaging practices.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies III: Computed Tomography