Related Experiment Video
Updated: Apr 20, 2026

Paper-Based Preconcentration and Isolation of Microvesicles and Exosomes
Published on: April 29, 2020
Evaporative concentration on a paper-based device to concentrate analytes in a biological fluid
Sharon Y Wong1, Mario Cabodi, Jason Rolland
1Department of Biomedical Engineering, Boston University , 44 Cummington Mall, Boston, Massachusetts 02215, United States.
Abstract:
We report the first demonstration of using heat on a paper device to rapidly concentrate a clinically relevant analyte of interest from a biological fluid. Our technology relies on the application of localized heat to a paper strip to evaporate off hundreds of microliters of liquid to concentrate the target analyte. This method can be used to enrich for a target analyte that is present at low concentrations within a biological fluid to enhance the sensitivity of downstream detection methods. We demonstrate our method by concentrating the tuberculosis-specific glycolipid, lipoarabinomannan (LAM), a promising urinary biomarker for the detection and diagnosis of tuberculosis. We show that the heat does not compromise the subsequent immunodetectability of LAM, and in 20 min, the tuberculosis biomarker was concentrated by nearly 20-fold in simulated urine. Our method requires only 500 mW of power, and sample flow is self-driven via capillary action. As such, our technology can be readily integrated into portable, battery-powered, instrument-free diagnostic devices intended for use in low-resource settings.
Related Concept Videos
Dialysis
Capillary Electrophoresis: Instrumentation