Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Cone-beam CT versus multi-slice CT systems for postoperative imaging of cochlear implantation--a phantom study on
Henricus J Theunisse1, Raoul M S Joemai, Thomas J J Maal
1*Department of Otorhinolaryngology and Head and Neck Surgery, Radboud university medical center and Hearing and Implants, Donders Institute for Brain, Cognition and Behaviour, Nijmegen, The Netherlands; †Department of Radiology, Leiden University Medical Centre, Leiden, The Netherlands; ‡Department of Oral and Maxillofacial Surgery, 3D Facial Imaging Research Group, Radboud university medical center, Nijmegen, The Netherlands; and §Department of Radiology, Radboud university medical center, Nijmegen, The Netherlands.
Hypothesis:
Image quality of low-dose multi-slice computed tomography (MSCT) after cochlear implantation is comparable to that of cone-beam computed tomography (CBCT).
Background:
CBCT has been described as a low-dose alternative with superior image quality to MSCT for postoperative cochlear implant (CI) imaging, but to our knowledge, no dose-matched comparisons of image quality have been published.
Materials And Methods:
Five human cochleae were implanted with CI electrodes and scanned on two CBCT and two MSCT systems. Four independent observers rated aspects of image quality on a five-point scale. CBCT scans were compared to clinical and dose-matched MSCT scans. Declining-dose MSCT protocols were compared to the clinical protocol. CT phantoms were used to determine effective dose and resolution for each acquisition protocol.
Results:
Effective dose of the CBCT protocols was 6 to 16% of the clinical MSCT dose. Visibility of cochlear inner and outer walls and overall image quality were positively correlated with radiation dose on MSCT and image quality was better with clinical MSCT than with CBCT protocols. In other comparisons, differences between systems were found, but a distinction between CBCT and MSCT could not be made.
Conclusion:
CBCT and dose-matched MSCT are both suitable for postoperative CI imaging. Selecting a CT system and radiation dose depends on which cochlear structures need to be visualized.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography