70 kVp computed tomography pulmonary angiography: potential for reduction of iodine load and radiation dose
Julian L Wichmann1, Xiaohan Hu, Josef M Kerl
1Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt am Main, Germany.
Purpose:
The purpose of the study was to evaluate 70 kVp dual-source computed tomography pulmonary angiography (CTPA) with reduced iodine load in comparison with single-source 70 and 100 kVp CTPA with standard iodine load regarding image quality and radiation dose.
Materials And Methods:
Three groups with 40 consecutive patients each underwent either standard single-source 100 kVp (120 mAs; group A), single-source 70 kVp (208 mAs; group B), or dual-source 70 kVp CTPA (416 mAs; group C). A volume of 70 mL of contrast material with 400 mg I/mL (groups A, B) or 300 mg I/mL (group C) was administered. Chest diameter, dose-length product, intravascular signal attenuation, image noise, signal to noise ratio (SNR), and contrast to noise ratio (CNR) were compared. Two observers rated subjective image quality regarding intravascular enhancement and image noise using 5-point scales.
Results:
Chest diameter and age were similar (P ≥ 0.28) for all groups. Compared with group A, the average dose-length product was 59% lower in group B (67.3 ± 11.8 vs. 164.7 ± 50.6 mGy cm, P<0.001) and similar between groups A and C (167.7 ± 41.2 mGy cm, P = 0.39). Average SNR and CNR were significantly higher for group C (21.5 ± 4.7 and 19.0 ± 4.5, respectively) compared with groups A (18.3 ± 3.5 and 15.8 ± 3.4, respectively) and B (17.3 ± 5.8 and 15.6 ± 5.5, respectively; all Ps ≤ 0.001). Subjective image quality ratings regarding enhancement and noise were highest for group C (1.73 ± 0.62 and 2.03 ± 0.66, respectively).
Conclusions:
Compared with standard 100 kVp CTPA, single-source 70 kVp CTPA allows for significant radiation dose savings with comparable SNR and CNR, whereas dual-source 70 kVp CTPA results in a superior objective image quality albeit a reduction of iodine concentration.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...