Chronic chest pain in a patient with an implantable cardioverter-defibrillator was linked to lead placement. Percutaneous lead extraction and repositioning successfully resolved the patient's symptoms.
Area of Science:
- Cardiology
- Medical Devices
Background:
- A 65-year-old male presented with 8-year history of chronic chest pain.
- The patient had a history of inducible ventricular tachycardia treated with a dual-chamber implantable cardioverter-defibrillator (ICD).
- Coronary angiography showed nonobstructive coronary artery disease and mild idiopathic nonischemic cardiomyopathy.
Observation:
- Chest radiography and transthoracic echocardiography revealed the ICD lead was embedded in the right ventricle's moderator band.
- Mild calcification was noted at the lead insertion site.
Findings:
- The patient underwent extraction of the dual-coil lead and implantation of a new single-coil right ventricular lead in the mid septum.
- Complete symptom relief was achieved post-procedure.
Implications:
- Right ventricular ICD lead placement in the moderator band can be a cause of chronic chest pain.
- Percutaneous lead extraction and repositioning can effectively alleviate such symptoms.
- This case highlights the importance of lead placement assessment in patients with unexplained chest pain.
Abstract:
A 65-year-old man was evaluated for chronic chest pain that had been present for 8 years after placement of a dual-chamber implantable cardioverter-defibrillator to treat inducible ventricular tachycardia. Previous coronary angiography had revealed nonobstructive coronary artery disease and a left ventricular ejection fraction of 0.45 to 0.50, consistent with mild idiopathic nonischemic cardiomyopathy. Evaluation with chest radiography and transthoracic echocardiography showed the implantable cardioverter-defibrillator lead to be embedded within the right ventricle at the moderator band, which had mild calcification. Treatment included extraction of the dual-coil lead and placement of a new single-coil right ventricular lead at the mid septum. The patient had complete relief of symptoms after the procedure. This case shows that chest pain can be associated with the placement of a right ventricular implantable cardioverter-defibrillator lead in the moderator band and that symptomatic relief can occur after percutaneous lead extraction and the implantation of a new right ventricular lead to the mid septal region.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Stenosis I: Introduction
Mitral Stenosis III: Medical Management
Mitral Stenosis IV: Nursing Management
Mitral Regurgitation IV: Nursing Management