Related Experiment Video
Updated: Apr 20, 2026

Guided Endodontics: Three-Dimensional Planning and Template-Aided Preparation of Endodontic Access Cavities
Published on: May 24, 2022
In vitro comparison of working length determination using three different electronic apex locators
Alper Kuştarci1, Dilara Arslan2, Demet Altunbaş3
1Department of Endodontics, School of Dentistry, Akdeniz University, Antalya, Turkey.
Background:
The aim of this study was to compare the accuracy of the apex-locating functions of DentaPort ZX, Raypex 5 and Endo Master electronic apex locators (EALs) in vitro.
Materials And Methods:
Thirty extracted human single-rooted teeth with mature apices were used for the study. The real working length (RWL) was established by subtracting 0.5 mm from the actual root canal length. All teeth were mounted in an alginate model that was especially developed to test the EALs and the teeth were then measured with each EAL. The results were compared with the corresponding RWL, which was subtracted from the electronically determined distance. Data were analyzed using a paired-samples t-test, a Chi-square test and a repeated measure analysis of variance evaluation at the 0.05 level of significance.
Results:
Statistical analysis showed that no significant difference was found among all EALs (P > 0.05).
Conclusion:
The accuracy of the EALs was evaluated and all of the devices showed an acceptable determination of electronic working length between the ranges of ±0.5 mm.
Related Concept Videos
Electronic Distance Measuring Instruments
Design Example: Measuring Distance Between Two Points with Obstructions
Differential Leveling
Common Leveling Mistakes and Errors
Adjusting a Traverse
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device