Multimodality Diagnosis of Mesenteric Ischemia
Published on: July 21, 2023
Early small-bowel ischemia: dual-energy CT improves conspicuity compared with conventional CT in a swine model
Theodora A Potretzke1, Christopher L Brace, Meghan G Lubner
1From the Departments of Radiology (T.A.P., C.L.B., M.G.L., L.A.S., B.J.W., F.T.L.), Biomedical Engineering (C.L.B.), and Medical Physics (C.L.B.), University of Wisconsin-Madison, E3/366 Clinical Science Center, 600 Highland Ave, Madison, WI 53792.
Purpose:
To compare dual-energy computed tomography (CT) with conventional CT for the detection of small-bowel ischemia in an experimental animal model.
Materials And Methods:
The study was approved by the animal care and use committee and was performed in accordance with the Guide for Care and Use of Laboratory Animals issued by the National Research Council. Ischemic bowel segments (n = 8) were created in swine (n = 4) by means of surgical occlusion of distal mesenteric arteries and veins. Contrast material-enhanced dual-energy CT and conventional single-energy CT (120 kVp) sequences were performed during the portal venous phase with a single-source fast-switching dual-energy CT scanner. Attenuation values and contrast-to-noise ratios of ischemic and perfused segments on iodine material-density, monospectral dual-energy CT (51 keV, 65 keV, and 70 keV), and conventional 120-kVp CT images were compared. Linear mixed-effects models were used for comparisons.
Results:
The attenuation difference between ischemic and perfused segments was significantly greater on dual-energy 51-keV CT images than on conventional 120-kVp CT images (mean difference, 91.7 HU vs 47.6 HU; P < .0001). Conspicuity of ischemic segments was significantly greater on dual-energy iodine material-density and 51-keV CT images than on 120-kVp CT images (mean contrast-to-noise ratios, 4.9, 4.3, and 2.1, respectively; P < .0001). Although attenuation differences on dual-energy 65- and 70-keV CT images were not significantly different from those on 120-kVp images (55.0 HU, 45.8 HU, and 47.6 HU, respectively; 65 keV vs 120 kVp, P = .15; 70 keV vs 120 kVp, P = .46), the contrast-to-noise ratio was greater for the 65- and 70-keV images than for the 120-kVp images (4.4, 4.1, and 2.1 respectively; P < .0005).
Conclusion:
Dual-energy CT significantly improved the conspicuity of the ischemic bowel compared with conventional CT by increasing attenuation differences between ischemic and perfused segments on low-kiloelectron volt and iodine material density images.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies for Cardiovascular System V: CT