Roll and pitch set-up errors during volumetric modulated arc delivery: can adapting gantry and collimator angles
Nienke A Hoffmans-Holtzer1, Daan Hoffmans, Max Dahele
1Department of Radiation Oncology, VU University Medical Center, De Boelelaan 1117, 1081, Amsterdam, The Netherlands.
Purpose:
The purpose of this work was to investigate whether adapting gantry and collimator angles can compensate for roll and pitch setup errors during volumetric modulated arc therapy (VMAT) delivery.
Methods:
Previously delivered clinical plans for locally advanced head-and-neck (H&N) cancer (n = 5), localized prostate cancer (n = 2), and whole brain with simultaneous integrated boost to 5 metastases (WB + 5M, n = 1) were used for this study. Known rigid rotations were introduced in the planning CT scans. To compensate for these, in-house software was used to adapt gantry and collimator angles in the plan. Doses to planning target volumes (PTV) and critical organs at risk (OAR) were calculated with and without compensation and compared with the original clinical plan. Measurements in the sagittal plane in a polystyrene phantom using radiochromic film were compared by gamma (γ) evaluation for 2 H&N cancer patients.
Results:
For H&N plans, the introduction of 2°-roll and 3°-pitch rotations reduced mean PTV coverage from 98.7 to 96.3%. This improved to 98.1% with gantry and collimator compensation. For prostate plans respective figures were 98.4, 97.5, and 98.4%. For WB + 5M, compensation worked less well, especially for smaller volumes and volumes farther from the isocenter. Mean comparative γ evaluation (3%, 1 mm) between original and pitched plans resulted in 86% γ < 1. The corrected plan restored the mean comparison to 96% γ < 1.
Conclusion:
Preliminary data suggest that adapting gantry and collimator angles is a promising way to correct roll and pitch set-up errors of < 3° during VMAT for H&N and prostate cancer.
Related Concept Videos
Common Leveling Mistakes and Errors
Differential Leveling
Errors in Taping
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Rotation with Constant Angular Acceleration - I
Using our intuition, we can begin to see how rotational quantities such as angular displacement, angular velocity, angular acceleration, and time are related to one another. For example, if a flywheel...
Rotation with Constant Angular Acceleration - II
The first...