Related Experiment Video
Updated: Apr 20, 2026

Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E sIgE
Published on: April 21, 2019
Results of an International Interlaboratory Trial to Determine Twelve Allergens Using Real-time PCR- and ELISA-based
René Köppel1, Jürg Rentsch2, Jürg Ruf3
1Official Food Control Authority of the Canton of Zurich, Zürich, Switzerland. rene.koeppel@klzh.ch.
To elucidate the capability of laboratories to determine allergen contents, an international interlaboratory trial was conducted using meat products spiked with 12 allergens. The measurement uncertainty was calculated independent of the applied method simulating realistic situations when comparing analysis certificates from different laboratories. The measurement uncertainty was revealed to be in the best cases +/-100%, in the worst cases quantification exhibited a measurement uncertainty of higher than 200% making quantitative analysis impossible. The measurement uncertainty seemed to depend on the analyte and assays used.
To elucidate the capability of laboratories to determine allergen contents, an international interlaboratory trial was conducted using meat products spiked with 12 allergens. The measurement uncertainty was calculated independent of the applied method simulating realistic situations when comparing analysis certificates from different laboratories. The measurement uncertainty was revealed to be in the best cases +/-100%, in the worst cases quantification exhibited a measurement uncertainty of higher than 200% making quantitative analysis impossible. The measurement uncertainty seemed to depend on the analyte and assays used.