Improved Performance of the Fourth-Generation FloTrac/Vigileo System for Tracking Cardiac Output Changes
Koichi Suehiro1, Katsuaki Tanaka1, Mika Mikawa1
1Department of Anesthesiology, Osaka City University Graduate School of Medicine, Osaka City, Osaka, Japan.
Objectives:
The aims of this study were to compare cardiac output (CO) measured by the new fourth-generation FloTrac™/Vigileo™ system (Version 4.00) (COFVS) with that measured by a pulmonary artery catheter (COREF), and to investigate the ability of COFVS to track CO changes induced by increased peripheral resistance.
Design:
Prospective study.
Setting:
University Hospital.
Participants:
Twenty-three patients undergoing cardiac surgery.
Interventions:
Phenylephrine (100 µg) was administered.
Measurements And Main Results:
Hemodynamic variables, including CO(REF) and CO(FVS), were measured before and after phenylephrine administration. Bland-Altman analysis was used to assess the discrepancy between CO(REF) and CO(FVS). Four-quadrant plot and polar-plot analyses were utilized to evaluate the trending ability of CO(FVS) against CO(REF) after phenylephrine boluses. One hundred thirty-six hemodynamic interventions were performed. The bias shown by the Bland-Altman analysis was-0.66 L/min, and the percentage error was 55.4%. The bias was significantly correlated with the systemic vascular resistance index (SVRI) before phenylephrine administration (p<0.001, r(2) = 0.420). The concordance rate determined by four-quadrant plot analysis and the angular concordance rate calculated using polar-plot analysis were 87.0% and 83.0%, respectively. Additionally, this trending ability was not affected by SVRI state.
Conclusions:
The trending ability of the new fourth-generation FloTrac™/Vigileo™ system after increased vasomotor tone was greatly improved compared with previous versions; however, the discrepancy of the new system in CO measurement was not clinically acceptable, as in previous versions. For clinical application in critically ill patients, this vasomotor tone-dependent disagreement must be decreased.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: