Implementation of a novel algorithm for generating synthetic CT images from magnetic resonance imaging data sets for
Joshua Kim1, Carri Glide-Hurst1, Anthony Doemer1
1Department of Radiation Oncology, Henry Ford Health System, Detroit, Michigan.
Purpose:
To describe and evaluate a method for generating synthetic computed tomography (synCT) images from magnetic resonance simulation (MR-SIM) data for accurate digitally reconstructed radiograph (DRR) generation and dose calculations in prostate cancer radiation therapy.
Methods And Materials:
A retrospective evaluation was performed in 9 prostate cancer patients who had undergone MR-SIM in addition to CT simulation (CT-SIM). MR-SIM data were used to generate synCT images by using a novel, voxel-based weighted summation approach. A subset of patients was used for weight optimization, and the number of patients to use during optimization was determined. Hounsfield unit (HU) differences between CT-SIM and synCT images were analyzed via mean absolute error (MAE). Original, CT-based treatment plans were mapped onto synCTs. DRRs were generated, and agreement between CT and synCT-generated DRRs was evaluated via Dice similarity coefficient (DSC). Dose was recalculated, and dose-volume metrics and gamma analysis were used to evaluate resulting treatment plans.
Results:
Full field-of-view synCT MAE across all patients was 74.3 ± 10.9 HU with differences from CTs of 2.0 ± 8.1 HU and 11.9 ± 46.7 HU for soft tissue structures (prostate, bladder, and rectum) and femoral bones, respectively. Calculated DSCs for anterior-posterior and lateral DRRs were 0.90 ± 0.04 and 0.92 ± 0.05, respectively. Differences in D99%, mean dose, and maximum dose to the clinical target volume from CT-SIM dose calculations were 0.75% ± 0.35%, 0.63% ± 0.34%, and 0.54% ± 0.33%, respectively, for synCT-generated plans. Gamma analysis (2%/2 mm dose difference/distance to agreement) revealed pass rates of 99.9% ± 0.1% (range, 99.7%-100%).
Conclusion:
Generated synCTs enabled accurate DRR generation and dose computation for prostate MR-only simulation. Dose recalculated on synCTs agreed well with original planning distributions. Further validation using a larger patient cohort is warranted.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:54MR Molecular Imaging of Prostate Cancer with a Small Molecular CLT1 Peptide Targeted Contrast Agent
Published on: September 3, 2013
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging