Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
Mucosal impedance discriminates GERD from non-GERD conditions
Fehmi Ates1, Elif Saritas Yuksel1, Tina Higginbotham1
1Division of Gastroenterology, Hepatology and Nutrition, Vanderbilt Medical Center, Nashville, Tennessee.
A new minimally invasive device measuring esophageal mucosal impedance (MI) shows promise for diagnosing gastroesophageal reflux disease (GERD). This method offers higher accuracy than traditional wireless pH monitoring for detecting chronic reflux.
Area of Science:
- Gastroenterology
- Medical Devices
- Diagnostic Technologies
Background:
- Current diagnostic tests for gastroesophageal reflux disease (GERD) lack accuracy in measuring reflux chronicity.
- A novel minimally invasive device assessing esophageal mucosal impedance (MI) has been developed as a potential marker for chronic reflux.
Purpose of the Study:
- To investigate MI patterns in patients with GERD and non-reflux conditions.
- To assess MI patterns before and after proton pump inhibitor (PPI) treatment.
- To compare the diagnostic performance of MI with wireless pH monitoring.
Main Methods:
- A prospective longitudinal study evaluated MI in 273 participants, including those with GERD, esophagitis, achalasia, eosinophilic esophagitis, and healthy controls.
- MI was measured at multiple esophageal locations relative to the squamocolumnar junction.
- MI measurements were compared before and after acid suppressive therapy and with wireless pH monitoring data.
Main Results:
- Significantly lower MI values were observed in patients with GERD and eosinophilic esophagitis compared to controls (P < .001).
- GERD patients exhibited a distinct MI pattern (low impedance near the squamocolumnar junction, increasing axially) that normalized with PPI treatment.
- MI demonstrated higher specificity (95%) and positive predictive value (96%) for detecting esophagitis than wireless pH monitoring (64% and 40%, respectively).
Conclusions:
- Esophageal mucosal impedance measurement is a highly specific and sensitive tool for diagnosing GERD.
- This minimally invasive technique offers superior diagnostic performance compared to wireless pH monitoring for GERD detection.
Related Concept Videos
Gastroesophageal Reflux Disease
Gastroesophageal Reflux Disease II: Clinical Features and Management
Clinical Manifestations
GERD presents itself in a multitude of ways, with symptoms varying from person to person. The hallmark symptoms are...
Mucosal Barrier of the Stomach
Within parietal cells, carbonic acid is first formed through the reaction of water and carbon dioxide. The dissociation of carbonic acid releases bicarbonate and hydrogen ions. The bicarbonate...
Gastroesophageal Reflux Disease I: Meaning and Pathophysiology
Peptic Ulcer Disease III: Clinical Manifestations and Diagnostic Studies
Few clinical manifestations differentiate gastric ulcers from duodenal ulcers. Distinctions in the location, timing, and pain relief are crucial for healthcare providers in differentiating between gastric and duodenal ulcers during clinical assessments.
Pathophysiology of Peptic Ulcer Disease: Mucosal Defense Factors