Related Experiment Video
Updated: Apr 20, 2026

09:10
Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021
2.4K
An integrated orthognathic surgery system for virtual planning and image-guided transfer without intermediate splint
Dae-Seung Kim1, Sang-Yoon Woo2, Hoon Joo Yang1
1Department of Oral and Maxillofacial Radiology, BK21, and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, South Korea.
Summary
This study introduces a 3D virtual planning and image-guided system for orthognathic surgery, improving accuracy in transferring surgical plans. The new system demonstrates comparable precision to conventional methods, enhancing surgical outcomes.
Area of Science:
- Oral and Maxillofacial Surgery
- Medical Imaging
- Computer-Aided Surgery
Background:
- Orthognathic surgery requires precise planning and execution for optimal patient outcomes.
- Conventional planning relies on 2D cephalometric radiography and physical models, which can be time-consuming and less accurate.
- Transferring virtual surgical plans to physical models often involves intermediate steps like splint fabrication.
Purpose of the Study:
- To develop and evaluate an integrated system for 3D virtual surgical planning and image-guided transfer in orthognathic surgery.
- To compare the accuracy of image-guided transfer with conventional model surgery.
- To assess the feasibility of directly transferring virtual plans without intermediate splints.
Main Methods:
- Development of an integrated system combining 3D virtual planning software with image-guided technology.
- Virtual planning of maxillary surgery for orthognathic patients.
- Direct transfer of virtual surgical plans to physical cast models using image guidance.
- Comparison of virtual surgery results with conventional paper and model surgery outcomes.
Main Results:
- The root mean square (RMS) difference between virtual surgery and conventional model surgery was 0.75 ± 0.51 mm.
- The RMS difference between virtual surgery and image-guided transfer results was 0.78 ± 0.52 mm.
- No significant difference was found between the accuracy of the image-guided system and conventional methods.
Conclusions:
- The integrated 3D virtual planning and image-guided orthognathic surgery system offers comparable accuracy to traditional methods.
- This system has the potential to replace physical model surgical planning.
- Direct image-guided transfer streamlines the process, eliminating the need for intermediate splints and improving efficiency.
Keywords:
Image-guided transferIntegrated surgery systemOrthognathic surgeryVirtual surgical planning