A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
In a difficult access scenario, supraglottic airway devices improve success and time to ventilation
Wolfgang A Wetsch1, Andreas Schneider, Robert Schier
1aDepartment of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne bInstitute of Medical Statistics, Informatics and Epidemiology, University of Cologne, Cologne, Germany.
Abstract:
The success of tracheal intubation (TI) is unacceptably low in unconventional positions. Supraglottic airway devices (SAD) have become an important alternative. An airway manikin was placed in a car, simulating an entrapped motor vehicle accident victim. The rescuer only had access through the driver's door. Participants were (n = 25) anaesthesiologists with experience in prehospital emergency medicine. They attempted to secure the airway by TI or an SAD (Ambu AuraOnce, iGel, laryngeal tube) in a random sequence. Performance was compared using the Wilcoxon signed-rank test. P values less than 0.05 were considered statistically significant. Fastest effective ventilation was achieved with iGel (11.5 ± 6.9 s, P < 0.001), followed by a laryngeal mask (15.1 ± 5.6 s, P < 0.001) and a laryngeal tube (17.6 ± 5.3 s, P < 0.001); TI was the slowest (42.8 ± 23.9 s, comparator). iGel (P < 0.001) and laryngeal mask (P = 0.01) also significantly outperformed the laryngeal tube. First ventilation was achieved significantly faster with SADs compared with TI. Success rates were also higher when using SADs.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...