Related Experiment Video
Updated: Apr 20, 2026

Use of Label-free Optical Biosensors to Detect Modulation of Potassium Channels by G-protein Coupled Receptors
Published on: February 10, 2014
Sensitive detection of C-reactive protein using optical fiber Bragg gratings
S Sridevi1, K S Vasu2, S Asokan3
1Department of Instrumentation and Applied Physics, Indian Institute of Science, Bangalore 560012, India.
Abstract:
An accurate and highly sensitive sensor platform has been demonstrated for the detection of C-reactive protein (CRP) using optical fiber Bragg gratings (FBGs). The CRP detection has been carried out by monitoring the shift in Bragg wavelength (ΔλB) of an etched FBG (eFBG) coated with an anti-CRP antibody (aCRP)-graphene oxide (GO) complex. The complex is characterized by Fourier transform infrared spectroscopy, X-ray photoelectron spectroscopy and atomic force microscopy. A limit of detection of 0.01mg/L has been achieved with a linear range of detection from 0.01mg/L to 100mg/L which includes clinical range of CRP. The eFBG sensor coated with only aCRP (without GO) show much less sensitivity than that of aCRP-GO complex coated eFBG. The eFBG sensors show high specificity to CRP even in the presence of other interfering factors such as urea, creatinine and glucose. The affinity constant of ∼1.1×10(10)M(-1) has been extracted from the data of normalized shift (ΔλB/λB) as a function of CRP concentration.