Related Experiment Video
Updated: Apr 20, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
A multi-institutional dosimetry audit of rotational intensity-modulated radiotherapy
Catharine H Clark1, Mohammad Hussein2, Yatman Tsang3
1Department of Medical Physics, Royal Surrey County Hospital, Guildford, UK; National Physical Laboratory, London, UK; NCRI Radiotherapy Trials Quality Assurance Group, UK.
Background:
Rotational IMRT (VMAT and Tomotherapy) has now been implemented in many radiotherapy centres. An audit to verify treatment planning system modelling and treatment delivery has been undertaken to ensure accurate clinical implementation.
Material And Methods:
34 institutions with 43 treatment delivery systems took part in the audit. A virtual phantom planning exercise (3DTPS test) and a clinical trial planning exercise were planned and independently measured in each institution using a phantom and array combination. Point dose differences and global gamma index (γ) were calculated in regions corresponding to PTVs and OARs.
Results:
Point dose differences gave a mean (±sd) of 0.1±2.6% and 0.2±2.0% for the 3DTPS test and clinical trial plans, respectively. 34/43 planning and delivery combinations achieved all measured planes with >95% pixels passing γ<1 at 3%/3mm and rose to 42/43 for clinical trial plans. A statistically significant difference in γ pass rates (p<0.01) was seen between planning systems where rotational IMRT modelling had been designed for the manufacturer's own treatment delivery system and those designed independently of rotational IMRT delivery.
Conclusions:
A dosimetry audit of rotational radiotherapy has shown that TPS modelling and delivery for rotational IMRT can achieve high accuracy of plan delivery.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...