Unexpected triggering of the dialysate blood leak detector by haemolysis
This case showed that it is possible for haemoglobin released by haemolysis in the extracorporeal circuit to pass through a high permeability (albumin retaining) dialyser membrane. In the incident described, the blood leak detector of the dialysis machine was activated before the patient became symptomatic. Haemolysis should be considered as a possible cause of blood leak alarms during dialysis with high flux membranes.
This case showed that it is possible for haemoglobin released by haemolysis in the extracorporeal circuit to pass through a high permeability (albumin retaining) dialyser membrane. In the incident described, the blood leak detector of the dialysis machine was activated before the patient became symptomatic. Haemolysis should be considered as a possible cause of blood leak alarms during dialysis with high flux membranes.
More Related Videos
11:12Ex Vivo Red Blood Cell Hemolysis Assay for the Evaluation of pH-responsive Endosomolytic Agents for Cytosolic Delivery of Biomacromolecular Drugs
Published on: March 9, 2013
08:21Author Spotlight: A Pharmacodissection Approach to Uncover Mechanisms in Cardiovascular Disease Risk Populations
Published on: July 21, 2023
Related Concept Videos
Hemodialysis II: Procedure and Complications
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Hypersensitivity Reactions: Cytolytic Reactions
Hemodialysis I: Introduction