Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Comparison of temporal to pulmonary artery temperature in febrile patients
Donna Furlong1, Diane L Carroll, Cynthia Finn
1Donna Furlong, RN, is staff nurse at the Cardiac Surgical Intensive Care Unit, Massachusetts General Hospital, Boston. Ms Donna has been a staff nurse for 34 years in this unit. Diane L. Carroll, PhD, RN, FAAN, is nurse researcher, Munn Center for Nursing Research, Massachusetts General Hospital, Boston. Dr Carroll has been the nurse researcher for the past 7 years and previously a clinical nurse specialist. Cynthia Finn, RN, is staff nurse at the Cardiac Surgical Intensive Care Unit, Massachusetts General Hospital, Boston. Ms Finn has been a staff nurse for 29 years in this unit. Diane Gay, RN, is staff nurse at the Cardiac Surgical Intensive Care Unit, Massachusetts General Hospital, Boston. Ms Gay has been a nurse in this unit for 31 years. Christine Gryglik, PhD(c), RN, is former clinical nurse specialist at the Cardiac Surgical Intensive Care Unit, Massachusetts General Hospital, Boston. Mr Gryglik is a doctoral candidate at the University of Massachusetts. Vivian Donahue, MSN, RN, is nurse director at the Cardiac Surgical Intensive Care Unit, Massachusetts General Hospital, Boston. Ms Donahue has been the nurse director in this unit for the past 4 years and previously a clinical nurse specialist.
Background:
As a routine part of clinical care, temperature measurement is a key indicator of illness. With the criterion standard of temperature measurement from the pulmonary artery catheter thermistor (PAT), which insertion of PAT carries significant risk to the patient, a noninvasive method that is accurate and precise is needed.
Objectives:
The purpose of this study was to measure the precision and accuracy of 2 commonly used methods of collecting body temperature: PAT considered the criterion standard and the temporal artery thermometer (TAT) in those patients with a temperature greater than 100.4°F.
Methods:
This is a repeated-measures design with each patient with a PAT in the intensive care unit acting as their own control to investigate the difference in PAT readings and readings from TAT in the core mode. Accuracy and precision were analyzed.
Results:
There were 60 subjects, 41 males and 19 females, with mean age of 60.8 years, and 97% (n = 58) were post-cardiac surgery. There was a statistically significant difference between PAT and TAT (101.0°F [SD, 0.5°F] vs 100.5°F [SD, 0.8°F]; bias, -0.49°F; P < .001). Differences in temperature between the 2 methods were clinically significant (ie, >0.9°F different) in 15 of 60 cases (25%). No TAT measurements were 0.9 F greater than the corresponding PAT measurement (0%; 95% confidence interval, 0%-6%).
Discussion:
These data demonstrate the accuracy of TAT when compared with PAT in those with temperatures of 100.4°F or greater. This study demonstrates that TAT set to core mode is accurate with a 0.5°F lower temperature than PAT. There was 25% in variability in precision of TAT.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Increased Body Temperature
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...