A system for remote dosimetry audit of 3D-CRT, IMRT and VMAT based on lithium formate dosimetry
Emelie Adolfsson1, Håkan Gustafsson2, Eva Lund3
1Department of Radiation Physics and Radiation Physics, Department of Medical and Health Sciences, Linköping University, Linköping, Sweden.
Abstract:
The aim of this work was to develop and test a remote end-to-end audit system using lithium formate EPR dosimeters. Four clinics were included in a pilot study, absorbed doses determined in the PTV agreed with TPS calculated doses within ±5% for 3D-CRT and ±7% (k=1) for IMRT/VMAT dose plans.