Transcatheter patent ductus arteriosus closure with echocardiographic guidance: can radiation exposure be reduced?

Ibrahim Cansaran Tanıdır1, Alper Güzeltaş2, Yakup Ergül2

  • 1Department of Pediatric Cardiology, İstanbul Mehmet Akif Ersoy Thoracic and Cardiovascular Surgery Center and Research Hospital, İstanbul, Turkey. cansaran@yahoo.com.

Insights

Reducing radiation exposure in pediatric cardiac interventions is crucial. This study shows transcatheter closure of patent ductus arteriosus (PDA) in children can be done with less radiation and contrast using echocardiography instead of aortography.

Area of Science:

  • Pediatric Cardiology
  • Interventional Radiology
  • Medical Imaging

Background:

  • Children are more sensitive to radiation than adults, making radiation exposure during cardiac catheterization a significant concern.
  • Transcatheter closure of patent ductus arteriosus (PDA) and other pediatric interventions raise concerns about radiation dose.

Purpose of the Study:

  • To demonstrate a method for transcatheter closure of PDA in children using reduced ionizing radiation.
  • To prove that radiation dose and contrast material can be decreased during pediatric PDA closure.

Main Methods:

  • Device selection was based on PDA morphology and diameter.
  • Procedures utilized Gianturco coils, Amplatzer Duct Occluders (ADO), Flipper coils, and Amplatzer vascular plugs.
  • Echocardiography and control aortography findings were analyzed.

Main Results:

  • Fluoroscopy time, radiation dose, and contrast volume were 12 ± 6.4 mins, 28.1 ± 14.7 cmGy/cm²/kg, and 4.2 ± 2.3 cc/kg, respectively.
  • Control aortography revealed residual shunts in 39.7% of patients; real-time echocardiography showed shunts in 9.5%.
  • Control aortography accounted for 13.3% of total radiation and 27.2% of total contrast.

Conclusions:

  • Performing echocardiography instead of a final control aortography after transcatheter PDA closure can reduce patient radiation and contrast exposure.
  • Echocardiographic evaluation is a viable alternative to aortography for assessing residual shunts post-procedure.
Abstract