Evaluation of a nasal cannula in noninvasive ventilation using a lung simulator
Narayan P Iyer1, Robert Chatburn2
1Division of Neonatology, Department of Pediatrics, MetroHealth Medical Center, Case Western Reserve University, Cleveland, Ohio.
Background:
Nasal noninvasive ventilation (NIV) is a common form of noninvasive respiratory mode used in newborn infants. A next-generation nasal cannula (Neotech RAM cannula) has recently been used to provide nasal NIV. The impact of the Neotech RAM cannula on the delivery of pressure needs to be studied.
Methods:
In this ex vivo experimental design, a lung simulator (IngMar ASL 5000, version 3.4) was programmed to model a neonate (∼1-3 kg of body weight) with normal-to-moderately affected lungs. We used a Covidien PB840 ventilator with NIV software activated to compensate for leaks. Nasal NIV was set at peak airway pressures of 15, 20, and 25 cm H2O and PEEP of 5, 6, and 7 cm H2O. Three sizes of the Neotech RAM cannula were used (prong outer diameters of 3.0, 3.5, and 4.0 mm). The nose was designed to keep the leak of the nares by the prongs to 30%. We also created a worst case leak (58% leak) by using the largest simulated nostril diameter with the smallest diameter Neotech RAM cannula prong. The outcome measure was the difference in pressures, referred to as leak effect, measured by the lung simulator relative to the set peak airway pressure and PEEP on the ventilator.
Results:
For the interface with 30% leak, leak effects of peak airway pressure during simulated nasal NIV were similar with all Neotech RAM cannula sizes, with 63-75% of peak airway pressure and 70-90% of PEEP being transmitted across the nasal interface. The worst case scenario produced a 92% leak effect in peak airway pressure and PEEP.
Conclusions:
When used with ≤ 30% leak, the Neotech RAM cannula interface results in clinically acceptable transmission of pressures. With > 50% leak, a clinically negligible amount of pressure is transmitted to the artificial lungs.
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: