Related Experiment Video
Updated: Apr 19, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Radiochromic film based dosimetry of image-guidance procedures on different radiotherapy modalities
Ahmad Nobah1, Saad Aldelaijan, Slobodan Devic
1Biomedical Physics Department, King Faisal Specialist Hospital & Research Centre, Riyadh, Saudi Arabia. anobah@kfshrc.edu.sa.
Abstract:
In this work we compare doses from imaging procedures performed on today's state-of-the-art integrated imaging systems using a reference radiochromic film dosimetry system. Skin dose and dose profile measurements from different imaging systems were performed using radiochromic films at different anatomical sites on a humanoid RANDO phantom. EBT3 film was used to measure imaging doses from a TomoTherapy MVCT system, while XRQA2 film was used for dose measurements from kilovoltage imaging systems (CBCT on 21eX and TrueBeam Varian linear accelerators and CyberKnife stereoscopic orthogonal imagers). Maximum measured imaging doses in cGy at head, thorax, and pelvis regions were respectively 0.50, 1.01, and 4.91 for CBCT on 21eX, 0.38, 0.84, and 3.15 for CBCT on TrueBeam, 4.33, 3.86, and 6.50 for CyberKnife imagers, and 3.84, 1.90, and 2.09 for TomoTherapy MVCT. In addition, we have shown how an improved calibration system of XRQA2 film can achieve dose uncertainty level of better than 2% for doses above 0.25 cGy. In addition to simulation-based studies in literature, this study provides the radiation oncology team with data necessary to aid in their decision about imaging frequency for image-guided radiation therapy protocols.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation