Steinar Lundemoen1, Venny Lise Kvalheim2, Øyvind Sverre Svendsen3
1Section for Cardiothoracic Surgery, Department of Heart Disease, Haukeland University Hospital, University of Bergen, Bergen, Norway.
Background:
Recent studies have focused on the use of fixed-rate intraaortic balloon pumping (IABP) during cardiopulmonary bypass (CPB) to achieve pulsatile flow. Because application of an IABP catheter may represent a functional obstruction within the descending aorta, we explored the effect of IABP-pulsed CPB-perfusion with special attention to perfusion above and below the IABP balloon.
Methods:
Sixteen animals received an IABP catheter that remained turned off position (NP group, n = 8) or was switched to an automatic mode of 80 beats/min during CPB (PP group, n = 8). Flow-data and pressure-data were obtained above and below the IABP balloon. Tissue perfusion was evaluated by microspheres.
Results:
IABP-pulsed CPB-perfusion, as assessed at 30 minutes on CPB, increased proximal mean aortic pressure (p < 0.05) and carotid artery blood flow (p < 0.001), but decreased distal mean aortic pressure (p < 0.001). The decrease of distal mean aortic pressure in the PP group was associated with a 75 % decrease (p < 0.001) of renal tissue perfusion. During nonpulsed perfusion the respective variables remained essentially unchanged compared with pre-CPB levels.
Conclusions:
Using IABP as a surrogate to achieve pulsatile perfusion during CPB contributes significantly to lowered aortic pressure in the distal portion of aorta and impaired tissue perfusion of the kidneys. The results are focusing on effects that may contribute to organ dysfunction and acute kidney injury. Consequently, assessment of perfusion pressure distal to the balloon should be addressed whenever IABP is used during CPB.