Related Experiment Video
Updated: Apr 19, 2026

Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro
Published on: December 3, 2018
Comparing velocity and fluid shear stress in a stenotic phantom with steady flow: phase-contrast MRI, particle image
1Department of Electrical and Computer Engineering, University of Louisville, 2210 S Brook St, Shumaker Research Building, Room 216, Louisville, KY, 40292, USA, imankhodarahmi@yahoo.com.
Object:
This study aims to validate phase-contrast magnetic resonance imaging (PC-MRI) measurements of a steady flow through a severe stenotic phantom using particle image velocimetry (PIV) and computational fluid dynamics (CFD).
Materials And Methods:
The study was performed in an axisymmetric 87 % area stenosis model using an inlet Reynolds number (Re) of 160, corresponding to a jet Re of 444. Velocity patterns and estimated fluid shear stresses from three modalities were analyzed and compared qualitatively and quantitatively.
Results:
Visual analysis via contour subtraction and Bland-Altman plots showed good agreement for flow velocities and less agreement for maximum shear stress (MSS). The Pearson's coefficients of correlation between PC-MRI and PIV were 0.97 for the velocity field and 0.82 for the MSS. The corresponding parameters between PC-MRI and CFD were 0.96 and 0.84, respectively.
Conclusion:
Findings indicate that PC-MRI can be implemented to estimate velocity flow fields and MSS; however, this method is not sufficiently accurate to quantify the MSS at regions of high shear rate.
More Related Videos
Related Concept Videos
Navier–Stokes Equations
Newtonian Fluid: Problem Solving
A velocity gradient forms within the fluid when a Newtonian fluid is placed between two parallel plates, with...
Typical Model Studies
Couette Flow
Velocity and Acceleration in Steady and Unsteady Flow
The acceleration can be generalized to any point in the flow, and expressed as components along three perpendicular directions, representing changes in velocity over...