Related Experiment Video
Updated: Apr 19, 2026

Multifocal Electroretinograms
Published on: December 4, 2011
ISCEV Standard for full-field clinical electroretinography (2015 update)
Daphne L McCulloch1, Michael F Marmor, Mitchell G Brigell
1School of Optometry and Vision Sciences, University of Waterloo, 200 University Avenue West, Waterloo, ON, N2L 3G1, Canada, daphne.mcculloch@uwaterloo.ca.
Abstract:
This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for full-field clinical electroretinography (ffERG or simply ERG). The parameters for Standard flash stimuli have been revised to accommodate a variety of light sources including gas discharge lamps and light emitting diodes. This ISCEV Standard for clinical ERGs specifies six responses based on the adaptation state of the eye and the flash strength: (1) Dark-adapted 0.01 ERG (rod ERG); (2) Dark-adapted 3 ERG (combined rod-cone standard flash ERG); (3) Dark-adapted 3 oscillatory potentials; (4) Dark-adapted 10 ERG (strong flash ERG); (5) Light-adapted 3 ERG (standard flash "cone" ERG); and (6) Light-adapted 30 Hz flicker ERG. ISCEV encourages the use of additional ERG protocols for testing beyond this minimum standard for clinical ERGs.