Development of an In Vitro Ocular Platform to Test Contact Lenses
Published on: April 6, 2016
Calculated in situ tear oxygen tension under hybrid contact lenses
Karen Lai Yin Lee1, Diana Phuong-Anh Nguyen, Timothy B Edrington
1Southern California College of Optometry, Marshall B. Ketchum University, Fullerton, CA.
Objective:
We calculated corneal surface oxygen tension under hybrid contact lenses (CLs) by extending existing models of CL oxygen diffusion to the circumstances of hybrid CL designs with tear vaults.
Methods:
Theoretical oxygen tensions at the corneal surface, if tear mixing and exchange are excluded, are calculated for hybrid CL (modern, high oxygen-permeable rigid center) designs with a single chamber corneal model using a computer software spreadsheet. Several specific in vivo instances of hybrid CL tear vaults are measured by both slitlamp biomicroscopy and optical coherence tomography (OCT).
Results:
No significant difference was found between central thickness of hybrid CL rigid portions measured by Vigor gauge and OCT. The amount of central tear vault thickness was found to be significantly greater, however, when interpolated from biomicroscope images (500-1,554 μm, simulating a poorly fit lens) than from either OCT images (0-91 μm) or direct measurement through OCT software (0-96 μm). Using all measurements, excluding any potential contributions from tear exchange or mixing, we predict corneal surface tear pO2 values under several hybrid CLs to range from zero under a lens with an excessive vault to 112 mm Hg under a lens with a minimal vault.
Conclusions:
Tear vaults beneath the gas permeable portion of the hybrid CLs of 100 μm or less resulted in acceptable corneal surface oxygen values. Where tear vaults are greater than about 100 μm, predicted anterior corneal surface pO2 values are less likely to be adequate for optimal corneal physiology. Any significant tear exchange, however, should increase these values. Although measuring tear thickness, we noted that the photographic biomicroscope and the OCT methods provided inconsistent results, and thus limited the reliability of our primary result.
More Related Videos
07:27Anterior Segment Organ Culture Platform for Tracking Open Globe Injuries and Therapeutic Performance
Published on: August 25, 2021
06:29Effect of Artificial Tear Formulations on the Metabolic Activity of Human Corneal Epithelial Cells after Exposure to Desiccation
Published on: May 2, 2020
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Transport in the Blood
Ophthalmic Drug Delivery Systems
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...