Orcun Gurbuz1, Abdulkadir Ercan2, Hakan Ozkan3
1Department of Cardiovascular Surgery, Balikesir University, School of Medicine, Balikesir, 10010, Turkey. gurbuzorcun@gmail.com.
Abstract:
An increasing number of reports reveal the safety and efficacy of catheter ablation for atrial fibrillation therapy even in patients with a history of prosthetic heart valve. This case report presents a 60 year old female patient who had a prosthetic mitral paravalvular leak as a complication of radiofrequency ablation. Surgery revealed a 15 mm hole and three broken knots in the posterolateral mitral annulus. This report demonstrates radiofrequency ablation for atrial fibrillation in patients with prosthetic mitral valve may cause a paravalvular leak likely as a result of thermal injury. Moreover, patient can be asymptomatic at first. The clinician should keep in mind such complication and the patient should be evaluated in terms of paravalvular leakage.