Optimisation of occupational radiological protection in image-guided interventions: potential impact of dose rate
1Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, SE-413 45 Gothenburg, Sweden. Department of Radiation Physics, Institute of Clinical Sciences, Sahlgrenska Academy at University of Gothenburg, SE-413 45 Gothenburg, Sweden.
Abstract:
The optimisation of occupational radiological protection is challenging and a variety of factors have to be considered. Physicians performing image-guided interventions are working in an environment with one of the highest radiation risk levels in healthcare. Appropriate knowledge about the radiation environment is a prerequisite for conducting the optimisation process. Information about the dose rate variation during the interventions could provide valuable input to this process. The overall purpose of this study was to explore the prerequisite and feasibility to measure dose rate in scattered radiation and to assess the usefulness of such data in the optimisation process.Using an active dosimeter system, the dose rate in the unshielded scattered radiation field was measured in a fixed point close to the patient undergoing an image-guided intervention. The measurements were performed with a time resolution of one second and the dose rate data was continuously timed in a data log. In two treatment rooms, data was collected during a 6 month time period, resulting in data from 380 image-guided interventions and vascular treatments in the abdomen, arms and legs. These procedures were categorised into eight types according to the purpose of the treatment and the anatomical region involved.The dose rate varied substantially between treatment types, both regarding the levels and the distribution during the procedure. The maximum dose rate for different types of interventions varied typically between 5 and 100 mSv h(-1), but substantially higher and lower dose rates were also registered. The average dose rate during a complete procedure was however substantially lower and varied typically between 0.05 and 1 mSv h(-1). An analysis of the distribution disclosed that for a large part of the treatment types, the major amount of the total accumulated dose for a procedure was delivered in less than 10% of the exposure time and in less than 1% of the total procedure time.The present study shows that systematic dose rate measurements are feasible. Such measurements can be used to give a general indication of the exposure level to the staff and could serve as a first risk assessment tool when introducing new treatment types or x-ray equipment in the clinic. For example, it could provide an indication for when detailed eye dose measurements are needed. It also gives input to risk management considerations and the development of efficient routines for other radiological protection measures.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging