Related Experiment Video
Updated: Apr 19, 2026

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Anti-signal recognition particle autoantibody ELISA validation and clinical associations
Rohit Aggarwal1, Chester V Oddis2, Danielle Goudeau2
1Department of Medicine, Division of Rheumatology and Clinical Immunology, University of Pittsburgh, Pittsburgh, PA, USA aggarwalr@upmc.edu.
Objective:
The aim of this study was to develop and validate a quantitative anti-signal recognition particle (SRP) autoantibody serum ELISA in patients with myositis and longitudinal association with myositis disease activity.
Methods:
We developed a serum ELISA using recombinant purified full-length human SRP coated on ELISA plates and a secondary antibody that bound human IgG to detect anti-SRP binding. Protein immunoprecipitation was used as the gold standard for the presence of anti-SRP. Serum samples from three groups were analysed: SRP(+) myositis subjects by immunoprecipitation, SRP(-) myositis subjects by immunoprecipitation and non-myositis controls. The ELISA's sensitivity, specificity, positive predictive value and negative predictive value were evaluated. Percentage agreement and test-retest reliability were assessed. Serial samples from seven SRP immunoprecipitation-positive subjects were also tested, along with serum muscle enzymes and manual muscle testing.
Results:
Using immunoprecipitation, we identified 26 SRP(+) myositis patients and 77 SRP(-) controls (including 38 patients with necrotizing myopathy). Non-myositis control patients included SLE (n = 4) and SSc (n = 7) patients. Anti-SRP positivity by ELISA showed strong agreement (97.1%) with immunoprecipitation (κ = 0.94). The sensitivity, specificity, positive predictive value, and negative predictive value of the anti-SRP ELISA were 88, 100, 100 and 96, respectively. The area under the curve was 0.94, and test-retest reliability was strong (r = 0.91, P < 0.001). Serial samples showed that anti-SRP levels paralleled changes in muscle enzymes and manual muscle testing.
Conclusion: