Activity determination of (41)Ar using efficiency extrapolation method and 4πβ(PS)-4πγ(NaI) coincidence system
Feng Xie1, Xuesong Li1, Yihua Dai1
1Northwest Institute of Nuclear Technology, PO Box 69-14, Xi'an 710024, PR China.
Abstract:
Noble gas (41)Ar was measured with a 4πβ-4πγ coincidence system, in which gamma- and beta-rays were respectively detected with a well-type NaI(Tl) and plastic scintillator (PS) detector. The activity of (41)Ar was determined from an efficiency extrapolation method, in which the beta detector efficiency was varied by electronic discrimination using the software developed under Visual basic. In addition, high resolution gamma spectroscopy with HPGe detector was also used for activity determination of (41)Ar, and the result was satisfactory in agreement with that obtain by the efficiency extrapolation method. This work demonstrated that the activity of (41)Ar can be accurately measured by efficiency extrapolation method.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Mass Analyzers: Common Types
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Overview
Other Nuclides: 31P, 19F, 15N NMR
While fluorine-19 and phosphorous-31 have high natural abundances (100%) and positive gyromagnetic ratios, nitrogen-15 has a low natural abundance and a negative gyromagnetic ratio. However, nitrogen-15 is still preferred over nitrogen-14 (which has a...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....