Background:
Monitoring lacosamide (LCM) helps optimize therapeutic dosing in some clinical settings. We developed a novel LC-MS/MS method for measuring serum LCM and O-desmethyl lacosamide (ODL).
Results/Discussion:
The sample preparation was protein precipitation with methanol. The total CV was less than 4.7%. Calibration range was 0.95-30.29 µg/ml and 0.95-30.41 µg/ml while the expanded linear range was 0.41-47.49 µg/ml and 0.34-48.17 µg/ml for LCM and ODL, respectively with analytical accuracy of 87.2-106.0%. 45 random serum samples collected from 23 individuals on either 200 mg/day or 400 mg/day of LCM therapy showed concentrations of 2.2-19.8 µg/ml for LCM and up to 2.5 µg/ml for ODL.
Conclusion:
A simple and sensitive LC-MS/MS assay was validated for quantification of LCM and ODL in human serum.
More Related Videos
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
09:58Spatial Quantification of Drugs in Pulmonary Tuberculosis Lesions by Laser Capture Microdissection Liquid Chromatography Mass Spectrometry LCM-LC/MS
Published on: April 18, 2018
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
High-Performance Liquid Chromatography: Types of Detectors