Related Experiment Video
Updated: Apr 19, 2026

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Spatial and time-course thermodynamics during pulmonary vein isolation using the second-generation cryoballoon in a
Mitsuru Takami1, Juna Misiri1, H Immo Lehmann1
1From the Mayo Clinic/St. Marys Campus, Rochester, MN.
Background:
Thermodynamics in the left atrium-pulmonary vein (PV) junction, phrenic nerve, and esophagus during PV isolation (PVI) using the second-generation cryoballoon are not known.
Methods And Results:
Twenty dogs underwent PVI using second-generation cryoballoon. Ablations were performed for ≤2 deliveries based on PVI without a bonus freeze. Inner balloon, balloon surface, and tissue temperatures were monitored during cryoablation. The tissue thermocouples were placed on the epicardial surface of the left atrium-PV junction, as well as on the phrenic nerve and within the esophagus. A total of 259 cryoballoon and 229 tissue tissue thermocouples profiles during 53 cryoablations of 40 PVs were analyzed. Acutely, PVI was achieved in 36 of 40 PVs (90%). Conductive tissue cooling spread radially from the balloon-left atrium-PV contact point. The lowest tissue temperatures were dependent on the distance of the tissue thermocouples to the balloon surface (r=0.85; P<0.001). In addition, blood flow leaks around the balloon had a warming effect on the balloon and tissue temperature profiles. Chronic isolation (mean, 48±16 days) was achieved in 27 of 36 PVs (75%). In 8 of 9 acutely isolated but with chronic reconnection PVs, the blood flow leak location was concordant with chronic reconnection gap. Although only 1 esophageal ulcerated lesion was observed, neither phrenic nerve palsy nor severe PV stenosis was seen in any dogs.
Conclusions:
Variance in tissue thermodynamics during cryothermal ablation depends on the distance from balloon and peri-balloon blood flow leaks. This information may be useful for successful PVI without severe complications.
More Related Videos
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015